Difference between revisions of "Documentation/Nightly/Modules/DoseComparison"
(Updated screenshots) |
|||
| Line 30: | Line 30: | ||
{|align="center" | {|align="center" | ||
| − | |[[File: | + | |[[File:20160121_DoseComparisonUI.png|thumb|499px|Dose Comparison module UI]] |
|[[File:SlicerRT_0.11_Gamma.png|thumb|720px|Example of the result gamma volume]] | |[[File:SlicerRT_0.11_Gamma.png|thumb|720px|Example of the result gamma volume]] | ||
|} | |} | ||
| Line 40: | Line 40: | ||
<!-- ---------------------------- --> | <!-- ---------------------------- --> | ||
{{documentation/{{documentation/version}}/module-section|Panels and their use}} | {{documentation/{{documentation/version}}/module-section|Panels and their use}} | ||
| − | * | + | * Reference dose volume: Dose distribution used as reference |
| − | * | + | * Evaluated dose volume: Dose distribution used as evaluation/compare |
| − | * | + | * Mask structure: Segmentation and contained segment whithin which the gamma index is calculated, but not outside (it will be 0) |
| + | * Distance-to-agreement (DTA): Mark dose voxel as failed if closest dose voxel with same value found in distance higher than this threshold, see Low paper below | ||
| + | * Dose difference criteria: Mark dose voxel as failed if dose difference is equal or greater than this threshold, see Low paper below | ||
| + | ** Maximum value / Custom: Reference dose value for threshold can be either the maximum dose in the dose map, or a custom cGy value | ||
| + | * Do not calculate gamma values for voxels below (analysis threshold): self-explanatory | ||
| + | * Use linear interpolation: self-explanatory | ||
| + | * Upper bound for gamma values: If gamma value would be higher than this value, the value will be set instead | ||
| + | * Dose threshold on reference only | ||
| + | * Gamma volume: Output volume to contain gamma map. Needs to be created using the combobox. Name will be automatically filled from the names of the input volumes | ||
<!-- | <!-- | ||
{{documentation/{{documentation/version}}/module-parametersdescription}} | {{documentation/{{documentation/version}}/module-parametersdescription}} | ||
| Line 49: | Line 57: | ||
<!-- ---------------------------- --> | <!-- ---------------------------- --> | ||
{{documentation/{{documentation/version}}/module-section|Similar Modules}} | {{documentation/{{documentation/version}}/module-section|Similar Modules}} | ||
| − | [http://www.slicer.org/slicerWiki/index.php/Documentation/Nightly/Modules/ | + | [http://www.slicer.org/slicerWiki/index.php/Documentation/Nightly/Modules/SegmentComparison Segment Comparison] |
<!-- ---------------------------- --> | <!-- ---------------------------- --> | ||
Revision as of 17:38, 21 January 2016
Home < Documentation < Nightly < Modules < DoseComparison
|
For the latest Slicer documentation, visit the read-the-docs. |
Introduction and Acknowledgements
|
This work is part of the SparKit project, funded by An Applied Cancer Research Unit of Cancer Care Ontario with funds provided by the Ministry of Health and Long-Term Care and the Ontario Consortium for Adaptive Interventions in Radiation Oncology (OCAIRO) to provide free, open-source toolset for radiotherapy and related image-guided interventions. | |||||||
|
Module Description
The DoseComparison module computes the difference between two co-registered dose volumes using the gamma dose distribution comparison method. This module uses the gamma Plastimatch algorithm internally.
Use Cases
Comparing dose distributions
Tutorials
Panels and their use
- Reference dose volume: Dose distribution used as reference
- Evaluated dose volume: Dose distribution used as evaluation/compare
- Mask structure: Segmentation and contained segment whithin which the gamma index is calculated, but not outside (it will be 0)
- Distance-to-agreement (DTA): Mark dose voxel as failed if closest dose voxel with same value found in distance higher than this threshold, see Low paper below
- Dose difference criteria: Mark dose voxel as failed if dose difference is equal or greater than this threshold, see Low paper below
- Maximum value / Custom: Reference dose value for threshold can be either the maximum dose in the dose map, or a custom cGy value
- Do not calculate gamma values for voxels below (analysis threshold): self-explanatory
- Use linear interpolation: self-explanatory
- Upper bound for gamma values: If gamma value would be higher than this value, the value will be set instead
- Dose threshold on reference only
- Gamma volume: Output volume to contain gamma map. Needs to be created using the combobox. Name will be automatically filled from the names of the input volumes