Introduction and Acknowledgements
Acknowledgments:
The UIowa QIN PET DICOM Extension was funded in part by Quantitative Imaging to Assess Response in Cancer Therapy Trials NIH grant U01-CA140206 and Quantitative Image Informatics for Cancer Research (QIICR) NIH grant U24 CA180918. License: Slicer License | |||||||
|
General Information
Module Type & Category
Type: CLI
Category: Quantification
Module Description
Program title | SUV Factor Calculator |
Program description | This command line interface module takes a PET DICOM series and creates a Real World Value Mapping (RWVM) file including 4 types of Standardized Uptake Value conversion factors (SUVbw, SUVlbm, SUVbsa, SUVibw). |
Program documentation-url | Source code on GitHub |
Usage
Use Cases, Examples
The patient size correction factors are summarized here, where weight is in kilograms and height is in centimeters.
- SUVbw:
- males & females: weight
- SUVlbm:
- males: 1.10 * weight - 128 * (weight/height)^2
- females: 1.07 * weight - 148 * (weight/height)^2
- SUVbsa:
- males & females: weight^0.425 * height^0.725 * 0.007184
- SUVibw:
- males: 48.0 + 1.06 * (height - 152)
- females: 45.5 + 0.91 * (height - 152)
Tutorials
Quick Tour of Features and Use
A list of panels in the interface, their features, what they mean, and how to use them.
|
Development
Section under construction. |
Information for Developers
References
Sugawara et al. "Reevaluation of the Standardized Uptake Value for FDG: Variation with Body Weight and Methods for Correction." Radiology. 1999.
http://pubs.rsna.org/doi/pdf/10.1148/radiology.213.2.r99nv37521